Brands     Vishay Measurements Group     Interactive Guide     Instruction Bulletins     B-129


Surface Preparation for Strain Gage Bonding

Surfaces Requiring Special Treatment

Dimensional or Mechanical Changes Due to Surface Preparation

For most materials, strain measurement results are usually not significantly changed by the surface preparation procedures described in this publication. Even with appreciable material removal, effects on the static mechanical properties of the test part are generally negligible compared to other error sources in the experiment. It should be understood, however, that removal of a plated or hardened surface layer, or of a surface layer with significant residual stresses, may noticeably affect the fatigue life or the wear characteristics of the part when operated under dynamic service conditions.



Page 20 of 23