Introduction
The need for calibration arises frequently in
the use of strain gage instrumentation. Periodic
calibration is required, of course, to assure the
accuracy and/or linearity of the instrument itself.
More often, calibration is necessary to scale the
instrument sensitivity (by adjusting gage factor or
gain) in order that the registered output
correspond conveniently and accurately to some
predetermined input. An example of the latter
situation occurs when a strain gage installation is
remote from the instrument, with measurable signal
attenuation due to leadwire resistance. In this
case, calibration is used to adjust the sensitivity
of the instrument so that it properly registers the
strain signal produced by the gage. Calibration is
also used to set the output of any auxiliary
indicating or recording device (oscillograph,
computer display, etc.) to a convenient scale
factor in terms of the applied strain.
Direct vs. Indirect
Calibration
There are basically two methods of calibration
available - direct and indirect. With direct
calibration, a precisely known mechanical input is
applied to the sensing element of the measurement
system, and the instrument output is compared to
this for verification or adjustment purposes. For
example, in the case of transducer instrumentation,
an accurately known load (pressure, torque,
displacement, etc.) is applied to the transducer,
and the instrument sensitivity is adjusted as
necessary to register the corresponding output.
Direct calibration of instrument systems in this
fashion is highly desirable, but is not ordinarily
feasible for the typical stress analysis laboratory
because of the special equipment and facilities
required for its valid implementation.
(
continued ...
)
|