(
... continued
)
Unless additional leadwires are used (
as demonstrated later
), simulating compressive strain by directly
shunting the remote active gage is usually difficult
to implement in practice. Since the purpose of shunt
calibration in this case is simply to scale the
instrument sensitivity as a means of compensating for
leadwire resistance, either upscale or downscale
calibration is equally suitable. Thus, it is
generally more convenient to shunt the adjacent dummy
arm as shown below, because this can be done right at
the instrument terminals. It should be apparent from
the figure that the calibration resistor must be
connected directly across the dummy to produce the
desired result.
Quarter-bridge circuit with active gage remote
from instrument.
Gage strain cannot be accurately simulated by
shunting from S- to P- (or from S- to P+). After
shunting the dummy with a calibration resistor
selected to simulate the appropriate strain, the
instrument sensitivity is adjusted to register the
same strain. At low strain levels, the result is
effectively the same as if the calibration had been
performed by shunting the active gage.
(
continued...
)
|